Manual of Weighing Controller A 810
Page
26
A.S.T. Angewandte SYSTEM-TECHNIK GmbH Dresden
b A S
I C
3.1.1.7. Power-on zero
setting “POZS”
This toggle decision enables or
disables the power-on zero setting
option. Range is selected at 3.1.1.3
(Lower Limit) and 3.1.1.4 (Upper Limit)
previous page.
default setting: off
3.1.1.8. Minimum Load “ML”
Set in terms of divisions in the range of
0 … 250
.
This value indicates the trigger for the
print out.
default setting: 20
Æ
If enabled and other settings are default
printing is possible when weight >2kg
is on load cell (20(ML)*300kg(FS)=2kg)
3000(DN)
3.1.1.9. Upper limit taring
range “ULtAR”
Set in terms of percentage of full scale
in the range of
0 … 100%
.
It indicates the weight above zero up to
which an enabled tare option is operating.
default setting: 100
3.1.1.10. Taring Mode “tAR_M”
This mode selection defines the action
after the tare command.
Set in terms of numbers in the range from
0, 1 or 2.
mode 0
tare always;
mode 1
tare only when stable;
mode 2
if stable
Æ
tare
not
stable
Æ
tare if stable
is reached
default setting: 2
TARE
-1
+1
cursor
left
t A R
M
0
/NET
GROSS
ENT
2
/NET
GROSS
0
SHIFT
>
FUNC
value
incr. decr.
cursor
right
TARE
P O Z S
0
TARE
/NET
GROSS
ENT
o F F P O Z S
o n P O Z S
/NET
GROSS
0
M L
0
/NET
GROSS
ENT
0 2 0
/NET
GROSS
0
SHIFT
>
FUNC
TARE
+1
value
incr. decr.
-1
-1
+1
cursor
left
U L t A R
0
/NET
GROSS
ENT
1 0 0
/NET
GROSS
0
SHIFT
>
FUNC
value
incr. decr.
cursor
right
TARE