Manual Weighing Controller A 810
Page
114
A.S.T. Angewandte SYSTEM-TECHNIK GmbH Dresden
Err 90 error block check sum
bad block check sum of transmission of
calibrated data from PC
Err 91
error block check sum in memory
approved for calibration
defective terminal
Err 92
memory approved for calibration 95 %
capacity utilization
warning, memory approved for calibration
nearly full
Err 93 memory approved for calibration overflow memory approved for calibration cannot accept
any more data
Err 99
invalid combination of functions
violation of commissioning rules
Err 101
Sequence error
when start signal is on, the Stop signal turns
on
Err 102
Sequence error
during weighing cycle, the Stop signal turns on
Err 103
Zero Alert
turns on, when Zero Range is exceeded
(3.1.1.3 and 3.1.1.4 at page 25)
Err 104
Sequence error
Start signal turns on when Near Zero
(3.2.3.2.13 at page 71) is exceeded (signal is
OFF) and Near Zero Confirmation is ON
(3.1.8.7 at page 45)
Err 105
Sequence error
Start signal turns on when SP1 is exceeded
and SP1 Confirmation is ON (3.1.8.8 at page
46)
Err 109
Sequence error
During sequence zycle input D18 or D19 (with
associated user function number “230” (s. at
page )) is interrupted; downtime 3.1.8.9 at
page 46 is to short
Err 110 external codeset selection
selected external codeset is greater than 9
Err 120 accumulation sum overflow
accumulation exceeds maximum value of
4.294.967.295
Err 129 unstable signal during calibration
Reduce vibration or set “STAN” 3.1.1.11 and
“STAR” 3.1.1.12 at page 27 to higher values
Err 196 no positive acknowledgement from
Profibus-module
no module integrated; reset A810
Err 197 no answer from Profibus-module
no module integrated; reset A810
Err 198 wrong PC-port is set for Profibus
set PC-port to “3”
Chart 2, States of Error