Manual of Weighing Controller A 810
Page
24
A.S.T. Angewandte SYSTEM-TECHNIK GmbH Dresden
b A S I C
S E
T
o F F
S E
T
o n
S E
T
3.1.1. Submenu “Basic”
This menu contains the basic adjustments affecting the ability of the scale to be approved for calibration.
The adjustment might be carried out before calibration.
3.1.1.1. Enable Setup “Set”
Alternative decision to enable access
to a reduced setup section via function
call 108.
When ON, only parameters that not affect
legal-for-trade settings are available.
When OFF function call has no affect.
default setting: off
Refer to 5.3 at page 93 for locked menus.
3.1.1.2. Select primary
measurement unit
“MU”
Select one of the units
KG, To, Gr, Lb, oz, N, KN and FU (free unit)
as primary unit.
This unit is used for scale settings and
calibration.
default setting: KG
The Weighing Controller A810 will be calibrated with this unit and after each restart this unit is shown.
Note:
Free Unit “FU” is scaled and calibrated as “KG”. Setting appropriate chars for “FU” is done in
“basic” menu 3.1.1.16 and 3.1.1.17 at page 28.
ENT
0
TARE
/NET
GROSS
ENT
/NET
GROSS
0
TARE
TARE
M U
0
/NET
GROSS
ENT
K G
/NET
GROSS
0
scroll between active
units
TARE
Summary of Contents for a810