3.1.7.4. Over-under-go-
compare mode “OUC-
MD”
This selection defines when Over and Under
signal is taken into account.
0: compare always
1: compare when judging input is ON
2: compare when complete output is ON
3: compare when complete output is ON
and weight will be hold during that time
default setting: 0
3.1.7.5. Upper-lower-limit-
compare mode “ULC-
MD”
This selection defines when Upper- and
Lower limit signal is taken into account.
0: compare always
1: compare when judging input is ON
default setting: 0
3.1.7.6. Compare inhibited
time “CITI”
This timer defines duration between
50…999ms after SP1, SP2 or CPS setpoint
is reached.
No comparison is done during that time.
default setting: 500ms
3.1.7.7. Judging time “JTI”
This timer defines duration between
0…9999ms after SP3/CPS compensation
setpoint is reached.
After
judging timer is expired the
complete output timer can start.
default setting: 1500
Note:
When “JTI” is set to 0, no judgement
is done. Refer to example 6.7 at page 105.
C I T I
0
/NET
GROSS
ENT
/NET
GROSS
0
FUNC
-1
+1
cursor
left
5 0 0
SHIFT
>
value
incr. decr.
cursor
right
J T I
0
/NET
GROSS
ENT
/NET
GROSS
0
FUNC
-1
+1
cursor
left
1 5 0 0
SHIFT
>
value
incr. decr.
cursor
right
TARE
O U C - M D
0
0
/NET
GROSS
ENT
/NET
GROSS
0
-1
+1
value
incr. decr.
TARE
TARE
TARE
U L C - M D
0
/NET
GROSS
0
ENT
/NET
GROSS
0
-1
+1
value
incr. decr.
Summary of Contents for a810
Page 2: ......