
APPENDIX E. COMMUNICATIONS
HF-2500A HIGH FREQUENCY WELD CONTROL
E-8
990-371
HOST
POWER SUPPLY
Increments
Range
Time Range
Increments
0.01
0.1 - 0.99
0.1 - 0.99 ms
0.01 ms
0.1
1.00 - 9.90
1.0 - 9.9 ms
0.1 ms
1.0
10.00 - 99.0
10.0 - 99.0 ms
1 ms
weld_energy
is the parameter that specifies the amount of weld energy. In the current
feedback mode,
weld_energy
is in unit of 0.001 kA. In the voltage feedback mode,
weld_energy
is in units of 0.001 V. In the power feedback mode,
weld_energy
is in units of
0.001 kW.
volt multiplier
is an index value for a table of resistance vs. a PID multiplier for voltage mode.
Note: Not used in versions where
RINDEXx
and
EINDEXx
are present.
resistance index
is an index value into a table of resistance vs. energy PID tables. If 0, then a
test pulse will occur on the next weld to determine the actual resistance (Note: customer
control of this value is not recommended)
energy index
is an index value into a PID energy vs. PID values table. (Note: customer control
of this value is not recommended)
Command
MONITOR {READ | SET}
<crlf>
[
parameter_name value
<crlf>]
<lf>
Control State
Any except while welding
Description
Provides control over the basic weld monitor settings of the Power Supply schedule. When
used with the "READ" keyword, the basic weld monitor settings of the currently loaded
schedule are returned (see MONITOR under Power Supply Originated Commands). When the
"SET" keyword is used, the host may set (change) the value of one or more of the parameters
of the basic weld monitor settings pertaining to the currently loaded schedule. The following
is a list of valid literal substitutions for the parameter_name and value variables:
MONTYPE1
{ KA | V | KW| R }
Monitor Type for pulse 1
UPPER1
{ limit_value }
Upper Limit for pulse 1
LOWER1
{ limit_value }
Lower Limit for pulse 1
ACTION1
{ none | STOP | INHIBIT | APC }
Out of Limit Action for
pulse 1
MONTYPE2
{ KA | V | KW| R }
Monitor Type for pulse 2
UPPER2
{ limit_value }
Upper Limit for pulse 2
LOWER2
{ limit_value }
Lower Limit for pulse 2
ACTION2
{ none | STOP }
Out of Limit Action for pulse 2
P1LDLY1
{delay_value}
Pulse 1 Lower Delay Start Time For
Lower Limit
P1LDLY2
{delay_value}
Pulse 1 Lower Delay End Time For
Lower Limit
P1UDLY1
{delay_value}
Pulse 1 Upper Delay Start Time For
Upper Limit
P1UDLY2
{delay_value}
Pulse 1 Upper Delay End Time For
Upper Limit
Summary of Contents for HF-2500A
Page 1: ...990 371 REV J HIGH FREQUENCY WELD CONTROL HF 2500A OPERATION MANUAL...
Page 88: ......
Page 100: ......
Page 118: ......
Page 144: ......