:SBUS<n> Commands
26
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
633
:SBUS<n>:CAN:TRIGger:PATTern:ID
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID <string>
<string> ::= "nn...n" where n ::= {0 | 1 | X | $}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:CAN:TRIGger:PATTern:ID command defines the CAN
identifier pattern resource according to the string parameter. This pattern,
along with the identifier mode (set by the
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE command), control the identifier
pattern searched for in each CAN message.
If the string parameter starts with "0x", it is a hexadecimal string made up
of hexadecimal and X (don't care) characters; otherwise, it is a binary
string made up of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:ID?
The :SBUS<n>:CAN:TRIGger:PATTern:ID? query returns the current settings
of the specified CAN identifier pattern resource in the 29- bit binary string
format.
Return Format
<string><NL> in 29-bit binary string format
Errors
•
See Also
•
"Introduction to :TRIGger Commands"
•
":SBUS<n>:CAN:TRIGger:PATTern:ID:MODE"
•
":SBUS<n>:CAN:TRIGger:PATTern:DATA"
N O T E
The ID pattern resource string is always 29 bits. Only 11 of these bits are used when the
:SBUS<n>:CAN:TRIGger:PATTern:ID:MODE is STANdard.
A string longer than 29 bits is truncated to 29 bits when setting the ID pattern resource.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...