:FUNCtion Commands
15
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
311
•
SQUare
•
ABSolute — Absolute Value
•
LOG — Common Logarithm
•
LN — Natural Logarithm
•
EXP — Exponential (e
x
)
•
TEN — Base 10 exponential (10
x
)
•
Filters:
•
LOWPass — Low pass filter — The FREQuency:LOWPass command
sets the - 3 dB cutoff frequency.
•
HIGHpass — High pass filter — The FREQuency:HIGHpass command
sets the - 3 dB cutoff frequency.
Filters operate on a single analog channel source or on a g(t) function
that is the addition, subtraction, or multiplication of analog channel
sources (specified by the GOFT commands).
•
Visualizations:
•
MAGNify — Operates on a single analog channel source or on a g(t)
function that is the addition, subtraction, or multiplication of analog
channel sources (specified by the GOFT commands).
•
TRENd — Measurement trend — Operates on a single analog channel
source. The TRENd:MEASurement command selects the measurement
whose trend you want to measure.
•
BTIMing — Chart logic bus timing — Operates on a bus made up of
digital channels. The BUS:YINcrement, BUS:YORigin, and BUS:YUNit
commands specify function values.
•
BSTate — Chart logic bus state — Operates on a bus made up of
digital channels. The BUS:YINcrement, BUS:YORigin, and BUS:YUNit
commands specify function values. The BUS:CLOCk and BUS:SLOPe
commands specify the clock source and edge.
The SOURce1, DISPlay, RANGe, and OFFSet (or REFerence) commands
apply to any function.
Reporting the Setup
Use :FUNCtion? to query setup information for the FUNCtion subsystem.
Return Format
The following is a sample response from the :FUNCtion? queries. In this
case, the query was issued following a *RST command.
:FUNC:OPER ADD;DISP 0;SOUR1 CHAN1;SOUR2 CHAN2;RANG +8.00E+00;OFFS
+0.0E+00;:FUNC:GOFT:OPER ADD;SOUR1 CHAN1;SOUR2 CHAN2
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...