![Agilent Technologies InfiniiVision 3000 X-Series Programmer'S Manual Download Page 402](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-x-series/infiniivision-3000-x-series_programmers-manual_2868684402.webp)
402
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
19
:MEASure Commands
:MEASure:FALLtime
Command Syntax
:MEASure:FALLtime [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:FALLtime command installs a screen measurement and
starts a fall- time measurement. For highest measurement accuracy, set the
sweep speed as fast as possible, while leaving the falling edge of the
waveform on the display. If the optional source parameter is specified, the
current source is modified.
Query Syntax
:MEASure:FALLtime? [<source>]
The :MEASure:FALLtime? query measures and outputs the fall time of the
displayed falling (negative- going) edge closest to the trigger reference. The
fall time is determined by measuring the time at the upper threshold of
the falling edge, then measuring the time at the lower threshold of the
falling edge, and calculating the fall time with the following formula:
fall time = time at lower threshold - time at upper threshold
Return Format
<value><NL>
<value> ::= time in seconds between the lower threshold and upper
threshold in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
N O T E
This command is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...