:MEASure Commands
19
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
407
:MEASure:OVERshoot
Command Syntax
:MEASure:OVERshoot [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:OVERshoot command installs a screen measurement and
starts an overshoot measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:OVERshoot? [<source>]
The :MEASure:OVERshoot? query measures and returns the overshoot of
the edge closest to the trigger reference, displayed on the screen. The
method used to determine overshoot is to make three different vertical
value measurements: Vtop, Vbase, and either Vmax or Vmin, depending on
whether the edge is rising or falling.
For a rising edge:
overshoot = ((Vmax- Vtop) / (Vtop- Vbase)) x 100
For a falling edge:
overshoot = ((Vbase- Vmin) / (Vtop- Vbase)) x 100
Vtop and Vbase are taken from the normal histogram of all waveform
vertical values. The extremum of Vmax or Vmin is taken from the
waveform interval right after the chosen edge, halfway to the next edge.
This more restricted definition is used instead of the normal one, because
it is conceivable that a signal may have more preshoot than overshoot, and
the normal extremum would then be dominated by the preshoot of the
following edge.
Return Format
<overshoot><NL>
<overshoot>::= the percent of the overshoot of the selected waveform in
NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
N O T E
This command is not available if the source is FFT (Fast Fourier Transform).
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...