:MEASure Commands
19
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
421
:MEASure:SOURce
Command Syntax
:MEASure:SOURce <source1>[,<source2>]
<source1>,<source2> ::= {<digital channels> | CHANnel<n> | FUNCtion
| MATH | WMEMory<r> | EXTernal}
<digital channels> ::= DIGital<d> for the MSO models
<n> ::= 1 to (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :MEASure:SOURce command sets the default sources for
measurements. The specified sources are used as the sources for the
MEASure subsystem commands if the sources are not explicitly set with
the command.
If a source is specified for any measurement, the current source is
changed to this new value.
If :MARKer:MODE is set to OFF or MANual, setting :MEASure:SOURce to
CHANnel<n>, FUNCtion, or MATH will also set :MARKer:X1Y1source to
source1 and :MARKer:X2Y2source to source2.
EXTernal is only a valid source for the counter measurement (and
<source1>).
Query Syntax
:MEASure:SOURce?
The :MEASure:SOURce? query returns the current source selections. If
source2 is not specified, the query returns "NONE" for source2. If all
channels are off, the query returns "NONE,NONE". Source2 only applies to
:MEASure:DELay and :MEASure:PHASe measurements.
Return Format
<source1>,<source2><NL>
<source1>,<source2> ::= {<digital channels> | CHAN<n> | FUNC | WMWM<r>
| EXT | NONE}
See Also:
•
"Introduction to :MEASure Commands"
•
•
•
•
N O T E
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
MATH.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...