![Agilent Technologies InfiniiVision 3000 X-Series Programmer'S Manual Download Page 183](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-x-series/infiniivision-3000-x-series_programmers-manual_2868684183.webp)
Root (:) Commands
6
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
183
:ACTivity
Command Syntax
:ACTivity
The :ACTivity command clears the cumulative edge variables for the next
activity query.
Query Syntax
:ACTivity?
The :ACTivity? query returns whether there has been activity (edges) on
the digital channels since the last query, and returns the current logic
levels.
Return Format
<edges>,<levels><NL>
<edges> ::= presence of edges (16-bit integer in NR1 format).
<levels> ::= logical highs or lows (16-bit integer in NR1 format).
bit 0 ::= DIGital 0
bit 15 ::= DIGital 15
See Also
•
"Introduction to Root (:) Commands"
•
•
N O T E
Because the :ACTivity? query returns edge activity since the last :ACTivity? query, you must
send this query twice before the edge activity result is valid.
N O T E
A bit = 0 (zero) in the <edges> result indicates that no edges were detected on that channel
(across the specified threshold voltage) since the last query.
A bit = 1 (one) in the <edges> result indicates that edges have been detected on that
channel (across the specified threshold voltage) since the last query.
(The threshold voltage must be set appropriately for the logic levels of the signals being
probed.)
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...