930
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31
:WAVeform Commands
Digital Channel Data (MSO models only)
The waveform record for digital channels is similar to that of analog
channels. The main difference is that the data points represent either
DIGital0,..,7 (POD1), DIGital8,..,15 (POD2), or any grouping of digital
channels (BUS1 or BUS2).
For digital channels, :WAVeform:UNSigned (see
) must be set to
ON.
Digital Channel POD Data Format
Data for digital channels is only available in groups of 8 bits (Pod1 = D0 -
D7, Pod2 = D8 - D15). The bytes are organized as:
If the :WAVeform:FORMat is WORD (see
) is WORD, every other
data byte will be 0. The setting of :WAVeform:BYTeorder (see
controls which byte is 0.
If a digital channel is not displayed, its bit value in the pod data byte is
not defined.
Digital Channel BUS Data Format
Digital channel BUS definitions can include any or all of the digital
channels. Therefore, data is always returned as 16- bit values. :BUS
commands (see
) are used to select the digital channels for a bus.
Reporting the Setup
The following is a sample response from the :WAVeform? query. In this
case, the query was issued following a *RST command.
:WAV:UNS 1;VIEW MAIN;BYT MSBF;FORM BYTE;POIN +1000;SOUR CHAN1;SOUR:SUBS
NONE
:WAVeform:SOURce
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
POD1
D7
D6
D5
D4
D3
D2
D1
D0
POD2
D15
D14
D13
D12
D11
D10
D9
D8
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...