:MEASure Commands
19
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
399
:MEASure:DELay
Command Syntax
:MEASure:DELay [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:DELay command places the instrument in the continuous
measurement mode and starts a delay measurement.
The measurement is taken as:
delay = t(<edge spec 2>) - t(<edge spec 1>)
where the <edge spec> definitions are set by the :MEASure:DEFine
command
Query Syntax
:MEASure:DELay? [<source1>][,<source2>]
The :MEASure:DELay? query measures and returns the delay between
source1 and source2. The delay measurement is made from the
user- defined slope and edge count of the signal connected to source1, to
the defined slope and edge count of the signal connected to source2. Delay
measurement slope and edge parameters are selected using the
:MEASure:DEFine command.
Also in the :MEASure:DEFine command, you can set upper, middle, and
lower threshold values.
It is the middle threshold value that is used when
performing the delay query.
The standard upper, middle, and lower
measurement thresholds are 90%, 50%, and 10% values between Vbase and
N O T E
The :MEASure:DELay command and the front-panel delay measurement differ from the
:MEASure:DELay? query.
The delay command or front-panel measurement run the delay measurement in auto-edge
select mode. In this mode, you can select the edge polarity, but the instrument will select
the edges that will make the best possible delay measurement. The source1 edge chosen
will be the edge that meets the polarity specified and is closest to the trigger reference
point. The source2 edge selected will be that edge of the specified polarity that gives the
first of the following criteria:
•
The smallest positive delay value that is less than source1 period.
•
The smallest negative delay that is less than source1 period.
•
The smallest absolute value of delay.
The :MEASure:DELay? query will make the measurement using the edges specified by the
:MEASure:DEFine command.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...