![Agilent Technologies InfiniiVision 3000 X-Series Programmer'S Manual Download Page 221](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-x-series/infiniivision-3000-x-series_programmers-manual_2868684221.webp)
:ACQuire Commands
7
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
221
:ACQuire:COMPlete
Command Syntax
:ACQuire:COMPlete <complete>
<complete> ::= 100; an integer in NR1 format
The :ACQuire:COMPlete command affects the operation of the :DIGitize
command. It specifies the minimum completion criteria for an acquisition.
The parameter determines the percentage of the time buckets that must be
"full" before an acquisition is considered complete. If :ACQuire:TYPE is
NORMal, it needs only one sample per time bucket for that time bucket to
be considered full.
The only legal value for the :COMPlete command is 100. All time buckets
must contain data for the acquisition to be considered complete.
Query Syntax
:ACQuire:COMPlete?
The :ACQuire:COMPlete? query returns the completion criteria (100) for
the currently selected mode.
Return Format
<completion_criteria><NL>
<completion_criteria> ::= 100; an integer in NR1 format
See Also
•
"Introduction to :ACQuire Commands"
•
•
•
Example Code
' AQUIRE_COMPLETE - Specifies the minimum completion criteria for
' an acquisition.
The parameter determines the percentage of time
' buckets needed to be "full" before an acquisition is considered
' to be complete.
myScope.WriteString ":ACQuire:COMPlete 100"
See complete example programs at:
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...