![Agilent Technologies InfiniiVision 3000 X-Series Programmer'S Manual Download Page 191](http://html1.mh-extra.com/html/agilent-technologies/infiniivision-3000-x-series/infiniivision-3000-x-series_programmers-manual_2868684191.webp)
Root (:) Commands
6
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
191
:DIGitize
Command Syntax
:DIGitize [<source>[,..,<source>]]
<source> ::= {CHANnel<n> | FUNCtion | MATH | SBUS{1 | 2}}
for the DSO models
<source> ::= {CHANnel<n> | DIGital<d> | POD{1 | 2}
| BUS{1 | 2} | FUNCtion | MATH | SBUS{1 | 2}}
for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The <source> parameter may be repeated up to 5 times.
The :DIGitize command is a specialized RUN command. It causes the
instrument to acquire waveforms according to the settings of the :ACQuire
commands subsystem. When the acquisition is complete, the instrument is
stopped.
If no argument is given, :DIGitize acquires the channels currently
displayed. If no channels are displayed, all channels are acquired.
See Also
•
"Introduction to Root (:) Commands"
•
•
•
•
•
, “:ACQuire Commands,” starting on page 219
•
, “:WAVeform Commands,” starting on page 923
N O T E
The :DIGitize command is only executed when the :TIMebase:MODE is MAIN or WINDow.
N O T E
To halt a :DIGitize in progress, use the device clear command.
N O T E
MATH is an alias for FUNCtion.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...