:FUNCtion Commands
15
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
321
:FUNCtion[:FFT]:WINDow
Command Syntax
:FUNCtion[:FFT]:WINDow <window>
<window> ::= {RECTangular | HANNing | FLATtop | BHARris}
The :FUNCtion[:FFT]:WINDow command allows the selection of four
different windowing transforms or operations for the FFT (Fast Fourier
Transform) function.
The FFT operation assumes that the time record repeats. Unless an
integral number of sampled waveform cycles exist in the record, a
discontinuity is created between the end of one record and the beginning
of the next. This discontinuity introduces additional frequency components
about the peaks into the spectrum. This is referred to as leakage. To
minimize leakage, windows that approach zero smoothly at the start and
end of the record are employed as filters to the FFTs. Each window is
useful for certain classes of input signals.
•
RECTangular — useful for transient signals, and signals where there are
an integral number of cycles in the time record.
•
HANNing — useful for frequency resolution and general purpose use. It
is good for resolving two frequencies that are close together, or for
making frequency measurements. This is the default window.
•
FLATtop — best for making accurate amplitude measurements of
frequency peaks.
•
BHARris (Blackman- Harris) — reduces time resolution compared to the
rectangular window, but it improves the capacity to detect smaller
impulses due to lower secondary lobes (provides minimal spectral
leakage).
Query Syntax
:FUNCtion[:FFT]:WINDow?
The :FUNCtion[:FFT]:WINDow? query returns the value of the window
selected for the FFT function.
Return Format
<window><NL>
<window> ::= {RECT | HANN | FLAT | BHAR}
See Also
•
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...