:TRIGger Commands
30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
889
:TRIGger:PATTern:QUALifier
Command Syntax
:TRIGger:PATTern:QUALifier <qualifier>
<qualifier> ::= {ENTered | GREaterthan | LESSthan | INRange | OUTRange
| TIMeout}
The :TRIGger:PATTern:QUALifier command qualifies when the trigger
occurs:
•
ENTered — when the pattern is entered.
•
LESSthan — when the pattern is present for less than a time value.
•
GREaterthan — when the pattern is present for greater than a time
value. The trigger occurs when the pattern exits (not when the
GREaterthan time value is exceeded).
•
TIMeout — when the pattern is present for greater than a time value. In
this case, the trigger occurs when the GREaterthan time value is
exceeded (not when the pattern exits).
•
INRange — when the pattern is present for a time within a range of
values.
•
OUTRange — when the pattern is present for a time outside of range of
values.
Pattern durations are evaluated using a timer. The timer starts on the last
edge that makes the pattern (logical AND) true. Except when the TIMeout
qualifier is selected, the trigger occurs on the first edge that makes the
pattern false, provided the time qualifier criteria has been met.
Set the GREaterthan qualifier value with the
:TRIGger:PATTern:GREaterthan command.
Set the LESSthan qualifier value with the :TRIGger:PATTern:LESSthan
command.
Set the INRange and OUTRange qualifier values with the
:TRIGger:PATTern:RANGe command.
Set the TIMeout qualifier value with the :TRIGger:PATTern:GREaterthan
command.
Query Syntax
:TRIGger:PATTern:QUALifier?
The :TRIGger:PATTern:QUALifier? query returns the trigger duration
qualifier.
Return Format
<qualifier><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...