:WAVeform Commands
31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
933
:WAVeform:DATA
Query Syntax
:WAVeform:DATA?
The :WAVeform:DATA query returns the binary block of sampled data
points transmitted using the IEEE 488.2 arbitrary block data format. The
binary data is formatted according to the settings of the
:WAVeform:UNSigned, :WAVeform:BYTeorder, :WAVeform:FORMat, and
:WAVeform:SOURce commands. The number of points returned is
controlled by the :WAVeform:POINts command.
In BYTE or WORD waveform formats, these data values have special
meaning:
•
0x00 or 0x0000 — Hole. Holes are locations where data has not yet been
acquired.
Another situation where there can be zeros in the data, incorrectly, is
when programming over telnet port 5024. Port 5024 provides a
command prompt and is intended for ASCII transfers. Use telnet port
5025 instead.
•
0x01 or 0x0001 — Clipped low. These are locations where the waveform
is clipped at the bottom of the oscilloscope display.
•
0xFF or 0xFFFF — Clipped high. These are locations where the
waveform is clipped at the top of the oscilloscope display.
Return Format
<binary block data><NL>
See Also
•
For a more detailed description of the data returned for different
acquisition types, see:
"Introduction to :WAVeform Commands"
•
•
•
•
•
•
•
Example Code
' QUERY_WAVE_DATA - Outputs waveform data that is stored in a buffer.
' Query the oscilloscope for the waveform data.
myScope.WriteString ":WAV:DATA?"
' READ_WAVE_DATA - The wave data consists of two parts: the header,
' and the actual waveform data followed by a new line (NL) character.
' The query data has the following format:
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...