:WAVeform Commands
31
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
935
:WAVeform:FORMat
Command Syntax
:WAVeform:FORMat <value>
<value> ::= {WORD | BYTE | ASCii}
The :WAVeform:FORMat command sets the data transmission mode for
waveform data points. This command controls how the data is formatted
when sent from the oscilloscope.
•
ASCii formatted data converts the internal integer data values to real
Y- axis values. Values are transferred as ASCii digits in floating point
notation, separated by commas.
ASCII formatted data is transferred ASCii text.
•
WORD formatted data transfers 16- bit data as two bytes. The
:WAVeform:BYTeorder command can be used to specify whether the
upper or lower byte is transmitted first. The default (no command sent)
is that the upper byte transmitted first.
•
BYTE formatted data is transferred as 8- bit bytes.
When the :WAVeform:SOURce is the serial decode bus (SBUS1 or SBUS2),
ASCii is the only waveform format allowed.
When the :WAVeform:SOURce is one of the digital channel buses (BUS1 or
BUS2), ASCii and WORD are the only waveform formats allowed.
Query Syntax
:WAVeform:FORMat?
The :WAVeform:FORMat query returns the current output format for the
transfer of waveform data.
Return Format
<value><NL>
<value> ::= {WORD | BYTE | ASC}
See Also
•
"Introduction to :WAVeform Commands"
•
•
•
•
Example Code
•
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...