:MTESt Commands
21
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
495
:MTESt:SCALe:X1
Command Syntax
:MTESt:SCALe:X1 <x1_value>
<x1_value> ::= X1 value in NR3 format
The :MTESt:SCALe:X1 command defines where X=0 in the base coordinate
system used for mask testing. The other X- coordinate is defined by the
:MTESt:SCALe:XDELta command. Once the X1 and XDELta coordinates are
set, all X values of vertices in the mask regions are defined with respect
to this value, according to the equation:
X = (X *
Δ
X) + X1
Thus, if you set X1 to 100 ms, and XDELta to 100 ms, an X value of 0.100
is a vertex at 110 ms.
The oscilloscope uses this equation to normalize vertices. This simplifies
reprogramming to handle different data rates. For example, if you halve
the period of the waveform of interest, you need only to adjust the
XDELta value to set up the mask for the new waveform.
The X1 value is a time value specifying the location of the X1 coordinate,
which will then be treated as X=0 for mask regions coordinates.
Query Syntax
:MTESt:SCALe:X1?
The :MTESt:SCALe:X1? query returns the current X1 coordinate setting.
Return Format
<x1_value><NL>
<x1_value> ::= X1 value in NR3 format
See Also
•
"Introduction to :MTESt Commands"
•
•
•
•
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...