220
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
7
:ACQuire Commands
The :ACQuire:TYPE NORMal command sets the oscilloscope in the normal
acquisition mode. For the majority of user models and signals, NORMal
mode yields the best oscilloscope picture of the waveform.
Averaging
The :ACQuire:TYPE AVERage command sets the oscilloscope in the
averaging mode. You can set the count by sending the :ACQuire:COUNt
command followed by the number of averages. In this mode, the value for
averages is an integer from 2 to 65536. The COUNt value determines the
number of averages that must be acquired.
High-Resolution
The :ACQuire:TYPE HRESolution command sets the oscilloscope in the
high- resolution mode (also known as
smoothing
). This mode is used to
reduce noise at slower sweep speeds where the digitizer samples faster
than needed to fill memory for the displayed time range. Instead of
decimating samples, they are averaged together to provide the value for
one display point. The slower the sweep speed, the greater the number of
samples that are averaged together for each display point.
Peak Detect
The :ACQuire:TYPE PEAK command sets the oscilloscope in the peak
detect mode. In this mode, :ACQuire:COUNt has no meaning.
Reporting the Setup
Use :ACQuire? to query setup information for the ACQuire subsystem.
Return Format
The following is a sample response from the :ACQuire? query. In this case,
the query was issued following a *RST command.
:ACQ:MODE RTIM;TYPE NORM;COMP 100;COUNT 8;SEGM:COUN 2
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...