:SBUS<n> Commands
26
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
631
:SBUS<n>:CAN:TRIGger:PATTern:DATA
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA <string>
<string> ::= "nn...n" where n ::= {0 | 1 | X | $}
<string ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $}
The :SBUS<n>:CAN:TRIGger:PATTern:DATA command defines the CAN data
pattern resource according to the string parameter. This pattern, along
with the data length (set by the
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command), control the data
pattern searched for in each CAN message.
If the string parameter starts with "0x", it is a hexadecimal string made up
of hexadecimal and X (don't care) characters; otherwise, it is a binary
string made up of 0, 1, and X (don't care) characters.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA?
The :SBUS<n>:CAN:TRIGger:PATTern:DATA? query returns the current
settings of the specified CAN data pattern resource in the binary string
format.
Return Format
<string><NL> in nondecimal format
Errors
•
See Also
•
"Introduction to :TRIGger Commands"
•
":SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth"
•
":SBUS<n>:CAN:TRIGger:PATTern:ID"
N O T E
If more bits are sent for <string> than specified by the
:SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth command, the most significant bits will be
truncated. If the data length is changed after the <string> is programmed, the added or
deleted bits will be added to or deleted from the least significant bits.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...