Repairing the Sampler
Power Supply
Removing the power supply
1
In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2
Switch off the sampler at the main power switch. Remove the power cable.
3
Remove the top cover, top plate, and foam (see
4
Remove the plate tray from the tray base.
5
Remove the transport assembly, (see
“Removing the sample transport"
6
Remove the sampling unit, (see
7
Remove the MTP main board, (see
8
Remove the fan, (see
9
Remove the bottom foam.
10
Remove the exhaust fan, (see
11
Remove the power switch light pipe from the coupler.
12
Remove the two screws which hold the power supply from the rear panel.
13
Lift the power supply out of the module.
14
Remove the coupler from the switch of the power supply.
When required
When defective
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required
0950-2528 (rev. G and higher only) Power supply
C A U T I O N
The MTP board is sensitive to electrostatic discharge. Always use the ESD strap (see
on page 107) when handling electronic boards.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......