Control Module Screens of the Well-plate Samplers
8
1100 Series WS MWS Reference Manual
267
EMF (Early Maintenance Feedback)
Use the
F1
-key (
EMF
) to set EMF limits. Choose
Setup limits
to select the
number of injection valve cycles or number of needle movements into the
needle seat at which you want to receive a warning. Changes must be
acknowledged with
F6
(
Done
).
If a set limit has been exceeded, a message box will pop up. If you press
Reset
,
the limits will be removed.
Ignore
will continue to keep the EMF flag set.
N O T E
An exceeded limit will not stop a sequence or run (information only, to plan maintenance
activities).