1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Leak Sensor Open
The leak sensor in the sampler has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes
•
Leak sensor not connected to the MTP board.
•
Defective leak sensor.
Suggested Actions
✔
Ensure the leak sensor is connected correctly.
✔
Exchange the leak sensor.