Repairing the Sampler
4
1100 Series WS MWS Reference Manual
113
10
Remove the silicon safety tube from the needle.
11
Replace the plate tray in the tray base. Re-install the side door and close the
front door.
12
In the user interface close the “Change Needle/Seat” function and exit the
maintenance mode.
N O T E
The needle must be centered in the needle pusher as all alignment by the well-plate
sampler is calculated from the needle pusher position.