144
1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Main Fan
Removing the main fan
1
In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2
Switch off the sampler at the main power switch. Remove the power cable.
3
Remove the top cover, top plate, and foam (see
4
Remove the plate tray from the tray base.
5
Remove the transport assembly, (see
“Removing the sample transport"
6
Disconnect the fan connector from the MTP main board.
7
Lift the fan, and carefully pull the fan out of the foam.
Installing the main fan
1
Install the new fan and connect the fan connector to the MTP main board.
2
Install the sample transport (see
“Installing the sample transport"
3
Install the top cover, top plate, and foam (see
4
Replace the plate tray in the tray base.
5
Turn ON the sampler, close the front door.
6
The system will reset automatically.
When required
When defective.
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required
3160-1017 Fan
C A U T I O N
The MTP board is sensitive to electrostatic discharge. Always use the ESD strap (see
on page 107) when handling electronic boards.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......