270
1100 Series WS MWS Reference Manual
8
Control Module Screens of the Well-plate Samplers
Use the
Esc
-key to receive
Views
on the
F5
-key. Choose
System
from the
pull-down menu. Use the
F3
-key (
Records
) to select the WPS. Use the
F5
-key
(
FW Update
) to enter the Update section. Select the a file from the PCMCIA
card (1367nnnn.DLB for the non-thermostatted WPS or 1368nnnn.DLB for the
thermostatted WPS, use 1377nnnn.DLB for the non-thermostatted micro WPS
or 1378nnnn.DLB for the thermostatted micro WPS) and press execute. When
the update has finished the update, press
F8
-key (
Transfer
) to return the
module into the normal mode (LED on module should stays yellow).
Use the
F5
-key (
FW Update
) to enter the Update section. Select the a file from
the PCMCIA card (1367nnnn.DLB for the non-thermostatted WPS or
1368nnnn.DLB for the thermostatted WPS, use 1377nnnn.DLB for the
non-thermostatted micro WPS or 1378nnnn.DLB for the thermostatted micro
WPS) and press execute. When the update has finished, press
F7
-key
(
Transfer
) to return the module into the normal mode (LED on module should
stay yellow).
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......