Repairing the Sampler
4
1100 Series WS MWS Reference Manual
123
Installing the
metering seal
1
Install the new metering seal. Press it firmly into position.
2
Reassemble the analytical head. Press the plunger assembly into the seal.
3
Put the two fixing screws in place and reinstall the analytical head to the
sampling unit.
4
Tighten alternately the two fixing screws with the 4
mm hex key
.
5
Connect the two capillaries to the analytical head (Use a 1/4 inch wrench if
you have a SST capillary or a 4 mm wrench if you have a fused silica
capillary).
6
Reinstall the corrugated leak tubing.
7
Close the front door.
8
In the user interface close the “Change Piston” function and exit the
maintenance mode.