Repairing the Sampler
4
1100 Series WS MWS Reference Manual
147
Installing the MTP main board
1
Install the new board. Ensure the ribbon cables are positioned in the slot in
the board.
2
Reconnect the connectors to the board.
3
Replace the connector screws in the back of the module.
4
Replace the M4 screw at the back of the module.
5
Install the sample transport, (see
“Installing the sample transport"
6
Install the top cover, top plate, and foam (see
7
Replace the plate tray in the tray base.
8
Turn ON the sampler, close the front door, the system will reset
automatically.
9
Check the firmware revision of the sampler. If the firmware revision is
older than the current firmware revision of the autosampler, update the
firmware, see
“Replacing the Autosampler Firmware"
10
Perform a “Sample Transport Self Alignment”.
N O T E
If a new sampler main board is installed, update the serial number information of the
sampler in the user interface, see procedure bellow.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......