1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Installing the injection valve assembly
1
Feed the cable of the new valve through to the SUD board.
2
Connect the valve cable to the connector labelled “valve assy”.
3
Install and tighten the three screws which fix the injection valve assembly.
4
Install the sampling unit, (see
“Installing the sampling unit"
5
Install the sample transport (see
“Installing the sample transport"
6
Install the top cover, top plate, and foam (see
7
Replace the plate tray in the tray base.
8
Turn ON the sampler, close the front door.
9
The system will reset automatically.
10
In the user interface perform a “Sample Transport Self Alignment”.