1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
7. Needle / Seat error
✔
Upgrade the firmware revision to A.04.14 or higher and the ChemStation
revision to A.08.04 or higher
✔
Check if the needle is installed (the sample transport comes without
needle)
✔
Check the needle position and correct alignment in the pusher
✔
Check if the seat is not blocked with any parts or material (crystals, glass)
✔
Perform an auto-alignment
✔
Check the connector from the needle carrier to the sample transport unit
✔
Check the sample transport connectors on the main board/sample
transport
✔
Check the connector on the bottom of the sample transport
✔
Change the needle (G1367-87200) and the seat (G1367-87101)
✔
Change the needle carrier assembly (G1367-60010)
✔
Change the sample transport unit (G1367-60019)
✔
Change the main board (G1367-69500)