264
1100 Series WS MWS Reference Manual
8
Control Module Screens of the Well-plate Samplers
System - Tests
Press
F3
from the
System
View and choose the WPS to access the WPS
Tests
screen.
The Tests screen gives you access to multiple tests and features such as the
automatic transport unit
Alignment
tool,maintenance procedures to
Change
parts on the WPS or additional test
Signals
to monitor various parameters
during the WPS operation.
Use the
F1
-key (
Align Trans
) to perform an automatic alignment of the WPS
transport unit.
WA R N I N G
Only the empty tray base should be installed, when performing the alignment. The
wellplates must be removed from the WPS tray base before performing the transport
unit alignment. If the wellplates are installed during the alignment procedure, the
transport unit might get jammed or become defective.