Repairing the Sampler
Installing the peristaltic pump motor
1
Install the new motor and tighten the two holding screws.
2
Take in the new motor and holder.
3
Connect the pump motor to the SUD board.
4
Re-assemble the motor holder and the pump base plate.
5
Re-install the air channel.
6
Assemble the peristaltic pump, (see
7
Install the sampling unit, (see
“Installing the sampling unit"
8
Install the sample transport assembly, (see
9
Install the top cover, top plate, and foam (see
10
Replace the plate tray in the tray base.
11
Turn ON the sampler, close the front door.
12
The system will reset automatically.
13
In the user interface perform a “Sample Transport Self Alignment”.