Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
65
Leak
A leak was detected in the sampler.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the MTP board.
Probable Causes
•
Loose fittings.
•
Broken capillary.
•
Leaking rotor seal or needle seat.
•
Defective metering seal.
Suggested Actions
✔
Ensure all fittings are tight.
✔
Exchange defective capillaries.
✔
Exchange the rotor seal or seat capillary.
✔
Exchange the metering seal.
N O T E
Make sure the leak sensor is thoroughly dry before restarting the well-plate sampler.
N O T E
Crossflow caused by a leaky rotor seal may cause a spill over of seat in bypass position of
valve.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......