Repairing the Sampler
4
1100 Series WS MWS Reference Manual
115
Installing the Needle Carrier Assembly
1
Install a new needle carrier (
G1367-60010)
on place
2
Install the three holding hex screws with the 2 mm hex key.
3
Plug in the flex board on the sample transport.
4
Pinch the holder clamp and reinsert the needle assembly into the needle
carrier.
5
Check the alignment of the needle in the needle pusher of the needle carrier
by viewing from several directions to see that it is aligned in the center of
the needle pusher.
6
Remove the silicon safety tube from the needle.
7
Replace the plate tray in the tray base.
8
Re-install the side door and close the front door.
9
In the user interface close the “Change Needle Carrier” function and exit
the maintenance mode. The instrument will reset.
When required
When the needle carrier is defect
Tools required
2 mm hex key 8710-2438 (supplied in accessory kit)
Parts required
G1367-60010 Needle Carrier assembly
N O T E
The needle must be centered in the needle pusher as all alignment by the well-plate
sampler is calculated from the needle pusher position.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......