Parts and Materials
5
1100 Series WS MWS Reference Manual
187
Table 47
Agilent 1100 to HP 1090 LC, HP 1040 DAD or Signal Distribution Module
Connector
01046-60202
Pin
HP 1090
Pin
Agilent 1100
Signal Name
Active
(TTL)
1
1 - White
Digital ground
NC
2 - Brown
Prepare run
Low
4
3 - Gray
Start
Low
7
4 - Blue
Shut down
Low
8
5 - Pink
Not connected
NC
6 - Yellow
Power on
High
3
7 - Red
Ready
High
6
8 - Green
Stop
Low
NC
9 - Black
Start request
Low
Table 48
Agilent 1100 to General Purpose
Connector
01046-60201
Pin
Universal
Pin
Agilent 1100
Signal Name
Active
(TTL)
1 - White
Digital ground
2 - Brown
Prepare run
Low
3 - Gray
Start
Low
4 - Blue
Shut down
Low
5 - Pink
Not connected
6 - Yellow
Power on
High
7 - Red
Ready
High
8 - Green
Stop
Low
9 - Black
Start request
Low
5 - Key
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......