Repairing the Sampler
4
1100 Series WS MWS Reference Manual
145
Exhaust Fan
Removing the exhaust fan
1
In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2
Switch off the sampler at the main power switch. Remove the power cable.
3
Remove the top cover, top plate, and foam (see
4
Remove the plate tray from the tray base.
5
Remove the transport assembly, (see
“Removing the sample transport"
6
Remove the MTP main board, (see
7
Remove the exhaust fan.
Installing the exhaust fan
1
Install the exhaust fan.
2
Install the MTP main board, (see
“Installing the MTP main board"
3
Connect the fan connector to the MTP main board.
4
Install the sample transport (see
“Installing the sample transport"
5
Install the top cover, top plate, and foam (see
6
Replace the plate tray in the tray base.
7
Turn ON the sampler, close the front door, the system will reset
automatically.
When required
When defective.
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required
3160-4097 Fan
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......