1100 Series WS MWS Reference Manual
4
Repairing the Sampler
MTP Main Board
Removing the MTP main board
1
In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2
Switch off the sampler at the main power switch. Remove the power cable.
3
Remove the top cover, top plate, and foam (see
4
Remove the plate tray from the tray base.
5
Remove the transport assembly, (see
“Removing the sample transport"
6
Remove the connector screws at the back of the module.
7
Remove the M4 screw at the back of the module.
8
Disconnect all connectors on the MTP board
9
Slide the MTP main board out of the sampler.
When required
When defective
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Parts required
G1367-66500 MTP main board
G1367-69500 Exchange MTP main board
C A U T I O N
The MTP board is sensitive to electrostatic discharge. Always use the ESD strap (see
on page 107) when handling electronic boards.
N O T E
This procedure requires reloading the sampler firmware and reprogramming the instrument
serial number.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......