30
1100 Series WS MWS Reference Manual
1
Installing the Sampler
Flow Connections to the Sampler
1
Connect the pump outlet capillary to port 1 of the injection valve.
2
Connect column-compartment inlet capillary to port 6 of the injection
valve.
3
Connect the corrugated waste tube to the seat adapter and the solvent
waste from the leak plane.
4
Ensure that the waste tube is positioned inside the leak channel.
5
Drive the tube from the peristaltic flush pump to the solvent bottle in the
solvent cabinet
6
Seat capillary: see recommendations in
Preparation
Sampler is installed in the LC system
Parts required
Parts form the accessory kits, see
WA R N I N G
When opening capillary or tube fittings, solvents may leak out. Please observe
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......