208
1100 Series WS MWS Reference Manual
6
Introduction to the Well-plate Sampler
Reflective light switches detect the presence and type of different trays. The
X-slide carries the antenna and electronics of a RF-sensor. This device has
multiple functions:
•
It allows to read and write information from a tag, located in the new tray.
•
It allows to increase the number of different trays.
•
It allows to read the revision and other data tags of the needle/sample
transport assembly and sampling unit.
Complex flex boards make the electrical connection to the various motors,
sensors and the MTP-board. The needle carrier has an integrated plate/vial
pusher with an additional linear encoder to sense vials and the presence of
plates.
The needle and the loop capillary are user-exchangeable.
The back of the needle/sample transport assembly has a cover to protect the
electronics from potential solvent vapor.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......