Installing the Sampler
1
1100 Series WS MWS Reference Manual
27
8
Re-install the left side door (take care of the magnet at the back).
9
Place the sampler on top of the thermostat. Make sure that the sampler is
correctly engaged in the thermostat locks.
10
Remove the tray and the plastic cover from the tray base, place the air
channel adapter into the sampler tray base. Make sure the adapter is fully
pressed down. This assures that the cold airstream from the thermostat is
correctly guided to the tray area of the well-plate sampler.
11
Re-install the tray.
12
Ensure the power switch on the front of the sampler is 0FF and the power
cables are disconnected.
13
Connect the cable between the well-plate sampler and the thermostat, see
“Connection at the rear of thermostatted Sampler"
Figure 7
Installation of Thermostat and Sampler
Air channel adapter
WA R N I N G
Do not disconnect or reconnect the well-plate sampler to thermostat cable when the
power cords are connected to either of the two modules. This will damage the
electronics of the modules.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......