92
1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Needle
Defective or dirty sensor on the sampling-unit flex board.
Sticking needle-arm assembly.
Defective needle-drive motor.
Mainpass
Valve not connected.
Defective injection valve.
Needle Up/Mainpass
Blockage in the sample loop or needle (no solvent flow).
Defective or dirty sensor on the sampling-unit flex board.
Sticking needle-arm assembly.
Defective needle-drive motor.
Valve not connected.
Defective injection valve.
Table 17
Step Failures (continued)
Step Function
Probable Failure Modes