16
1100 Series WS MWS Reference Manual
1
Installing the Sampler
Environment
Your autosampler will work within specifications at ambient temperatures
and relative humidity as described in
C A U T I O N
Do not store, ship or use your autosampler under conditions where temperature
fluctuations may cause condensation within the autosampler. Condensation will
damage the system electronics. If your autosampler was shipped in cold weather, leave
it in its box, and allow it to warm up slowly to room temperature to avoid condensation.
Table 1
Physical Specifications - sampler (G1367A / G1377A)
Type
Specification
Comments
Weight
15.5 kg (34.2 lbs)
Dimensions
(height × width × depth)
200 × 345 × 440 mm
(8 × 13.5 × 17 inches)
Line voltage
100 – 240 VAC,
±
10 % Wide-ranging
capability
Line frequency
50 or 60 Hz,
±
5 %
Power consumption (apparent power)
300 VA
Maximum
Power consumption (active power)
200 W
Maximum
Ambient operating temperature
4 to 55
∞
C (41 to 131
∞
F)
Ambient non-operating temperature
-40 to 70
∞
C (-4 to 158
∞
F)
Humidity
< 95 %, at 25 to 40
∞
C (77 to 104
∞
F)
Non-condensing
Operating Altitude
Up to 2000 m (6500 ft)
Non-operating altitude
Up to 4600 m (14950 ft)
For storing the
autosampler
Safety standards: IEC, CSA, UL
Installation Category II, Pollution Degree 2
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...