Repairing the Sampler
4
1100 Series WS MWS Reference Manual
125
10
Pinch the holder clamp, pull back and remove the needle assembly with the
loop capillary from the needle carrier.
11
Attach the 5/16 inch wrench to hold position at the needle assembly. Use
the 4 mm wrench to loosen the fitting of the loop capillary.
12
Pull the loop capillary out from the needle assembly.