Chapter 3, Operating Overview
Measurement Triggering Process
Measurement Triggering Process
Active Measurements
Only active measurements can be triggered and then queried for a measurement
result. Within the Test Set, measurements are differentiated between those that are
"active" and those that are "not active". The definition of an active measurement is
different for analog measurements than for CDMA measurements.
Definition of Active CDMA Measurement
All CDMA measurements are "active" if:
•
either of the following screens is displayed on the CRT of the Test Set
CDMA CELLULAR MOBILE RECEIVER TEST
CDMA CELLULAR MOBILE TRANSMITTER TEST
AND
•
the measurement is in the ON state
NOTE:
Only the Avg/Chan Power measurement is active on the CDMA CALL CONTROL screen.
Definition of Active Analog Measurement
An analog measurement is "active" if:
•
the field used to display the measurement result is located on the screen which is
currently displayed on the CRT of the Test Set
AND
•
the measurement is in the ON state