358
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\errorms.fb
Error Messages
Error
−
103
Invalid separator
The parser was expecting a separator and encountered an illegal character. For example,
the colon used to separate the FREQ and AMPL commands should be omitted in the
following command:
RFG:FREQ 850 MHZ:;AMPL -35
Error
−
104
Data type error
The parser recognized a data element different than one allowed. For example, numeric or
string data was expected but block data was encountered.
Error
−
105
GET not allowed
A Group Execute Trigger was received within a program message (see IEEE 488.2, 7.7).
Error
−
108
Parameter not allowed
More parameters were received than expected for the header. For example, the *ESE
common command only accepts one parameter; receiving *ESE 36,1 is not allowed.
Error
−
109
Missing parameter
Fewer parameters were received than required for the header. For example, the *ESE
common command requires one parameter; receiving *ESE is not allowed.
Error
−
110
Command header error
An error was detected in the header.
Error
−
111
Header separator error
A character which is not a legal header separator was encountered while parsing the
header.
Error
−
112
Program mnemonic too long
The header contains more than twelve characters (see IEEE 488.2,7.6.1.4).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com