71
Chapter 3, Operating Overview
Interaction Between Screens
Using your Test Set, duplicate the steps in
figure 7
to demonstrate how the
priority fields operate.
Figure 7
Example of How Priority Fields Work
Table 3
Priority RX TEST and TX TEST Fields
Priority Field
RX TEST
TX TEST
RF Gen Amplitude
Presets to
−
80 dBm (changeable)
Always Off
AFGen1 To
Presets to FM (changeable)
Always Audio Out
AF Anl In
Always Audio In
Presets to FM Demod (changeable)
Detector
Always RMS
Presets to Pk +
−
Max(changeable)
De-emphasis
Always Off
Presets to 750
µ
s (changeable)
AF Anl Measurement
Presets to SINAD (changeable)
Presets to Audio Freq (changeable)
-50
Amplitude
dBm
-50
Amplitude
dBm
-50
Amplitude
dBm
-
7 5
1. Access the RX TEST screen and set the
Amplitude to -50 dBm.
-50
Amplitude
dBm
-75
Amplitude
dBm
2. Access the RF GENERATOR screen and
set the Amplitude to -75 dBm.
3. Access the RX TEST screen. Notice that
the Amplitude changes back to -50 dBm.
2. Return to the RF GENERATOR screen.
The Amplitude is no longer where you set it
in step 2 (-75 dBm).
Since the RX TEST screen has priority control over this field, the RF GENERATOR screen’s
Amplitude setting changed when RX TEST was accessed.