Chapter 3, Operating Overview
Setting an RF Generator/Analyzer Frequency Offset
Setting an RF Generator/Analyzer Frequency Offset
You can set a fixed frequency offset between the RF Generator and the RF
Analyzer. At power-up and instrument preset this feature is ON with a
transmit/receive frequency offset of 45 MHz.
To Turn Off RF Frequency Offset
1. Access the CONFIGURE screen.
2. Position the cursor below the
RF Display
field and select
Freq
.
3. Set the
RF Offset
to
Off
.
GPIB Example
“CONF:OMOD ‘OFF’”
turns the RF frequency offset off
To Change the RF Frequency Offset
1. Access the CONFIGURE screen.
2. Position the cursor below the
RF Display
field and select
Freq
.
3. Set the
RF Offset
to
On
.
4. Enter an offset frequency (
(Gen)-(Anl)
).