
69
Chapter 3, Operating Overview
Setting RF Generator/Analyzer Level Offsets
Setting RF Generator/Analyzer Level Offsets
The Test Set relies on level offsets to correctly set up the input signal path
attenuators. The user must enter path loss figures on the forward and reverse
channel paths independently by accessing the CONFIGURE screen. The Output
Port and Input Port fields select which port is being used on the front panel, and
the RF Level Offset field turns the level offset on.
At power-up and instrument preset this feature is OFF with 0.0 dB loss entered for
all connector fields.
Refer to “Determining and Correcting for RF Path Loss” in Calibrating the Test
Set chapter of the Agilent E8285A Application Guide for detailed procedures.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com